Part of Anesthesia Workstation
Main Body : High-strength engineering plastic rack, light, beautiful and corrosion resistant
Scope of Application : Adult and Child
Gas Source : O2: 0.28~0.6MPa N2O: 0.28~0.6MPa
Flow Meter : O2: 0.05~1.0L/min 1.1~10L/min
N2O: 0.05~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply : 25~75L/min
Low oxygen pressure alarm : There will be sound alarm when the oxygen pressure < 0.2MPa
Vaporizer : It has the function of automatic compensation based on pressure, temperature, and flow rate. The regulation range of evaporator concentration is 0~5 vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
Respiratory circuit working mode: all-close, semi-close, semi-open APL≤12.5kpa
Respiratory Bellows : Bellows for adults, bellows for children tidal volume range:0~1,500 ml
Part of Ventilator
Display mode : 10.4”TFT color LCD screen display
Ventilation mode : IPPV, SIPPV, SIMV, MANUAL, VCV, PCV
Ventilation function : PEEP,SIGH, IRV, IP, standby
Ventilation parameters : Tidal volume20~1600 ml
Rate 2~99 bpm
SIMV rate 2~20 bpm
I:E4:1~1:8
Inspiratory trigger pressure : -10~20cmH2O
PEEP : 2~20cmH2O
Pressure Range : 5~60cmH2O
SIGH 1.5 times the inspiratory time (60-120 adjustable)
Inspiratory platform : 0~50%
Pressure control : 3~50cmH2O
Pressure slope : 1~10 gear
Parameters for ventilation monitoring tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway,pressure - time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure(Inhale tidal volume, Inspiratory platform, average pressure, autonomous inspiratory rate, oxygen concentration, battery capacity, flow volume loop , pressure volume loop
Security Alarm System
Oxygen concentration alarm : Upper limit setting range 21%~99%
Low limit setting range 18%~80%
Airway pressure alarm : Upper limit setting range 5~60 cmH2O
Low limit setting range 0~50 cmH2O
Minute ventilation volume alarm : Upper limit setting range 1~20 L/min
Low limit setting range 0~19L/min
Tidal volume : Upper limit setting range 100~1500ml
Low limit setting range0~1400ml
Respiratory rate : Upper limit setting range 10~99bpm
Low limit setting range0~50bpm
Sustained high-pressure alarm : It will give alarm when stress have consistently been higher than 2.5cm H2O
Suffocation alarm, Power alarm, Intubation off alarm, battery voltage alarm
Power : AC 220V 50Hz UPS, Storage battery
Mechanical Arms (End-tidal CO2 monitoring) (Specific interface of waste anesthesia gas)
*Items within brackets are optional.
We are Exporters and Suppliers of Anesthesia Workstation Silver at wholesale competitive prices from India.
Model No :
MAN APAW 101 Anesthesia Workstation Silver